Models for ETO Sterilizers
Product forms | manual door cabinet |
Power source | AC380V,50Hz |
Sterilizer size | (Length × width × height) |
Sterilizer volume | From 1 to 100 cubic meters |
Sterilizer interior material | 304 Anticorrosive stainless steel |
Outer carton material | Carbon steel |
Temperature during sterilization | 50℃ ±3℃ |
Working pressure | -60kpa |
Maximum anti-stress value | -150kpa |
Humidity during sterilization | 45%-75% |
Pressure sensor | The built-in pressure sensor automatically detects the pressure inside the cavity to ensure that the pressure in the cavity is accurate and effective. |
Ethylene oxide sterilization concentration | 1g/L Adjustable |
Leakage rate of sterilization chamber | The leakage rate is measured under the conditions of no-load and constant temperature in the sterilization chamber, the test pressure is -50kPa, and the measurement time is 1h. The average leakage rate within the measurement time is ≤0.16kPa / min. |
Sterilization time | 0-99 hours Adjustable |
Vacuum speed | 15-25 minutes to -60kpa |
Air replacement | Automatic ventilation, ventilation times and interval time can be set freely |
Automatic control system | One-key start to end, the whole process of automatic completion with printing records |
Display status | Display temperature, humidity, pressure, sterilization time, residual gas treatment time, frequency |
Printer | Record and print the sterilization information of the whole process for verification. |
Humidity sensor | The built-in humidity sensor automatically detects the humidity in the cavity to ensure that the humidity in the cavity is not less than 45% RH. |
Residual gas treatment device | The removal rate of the filter with a particle diameter ≥ 0.3 μm is not less than 99.5%. |
Temperature sensor | The built-in temperature sensor automatically detects the temperature in the cavity to ensure that the temperature in the cavity is maintained within an accurate range of 50 ° C ± 3 ° C. |
Humidification system | Pulse humidification under vacuum to ensure that the humidity in the cavity is uniform and controllable; the humidity in the sterilization chamber should be controlled within the range of relative humidity 45% -80%; |
alarm system | High temperature alarm system, ultra-low pressure alarm system; When the sterilization temperature of sterilization chamber exceeds 59℃, the alarm shall be audible. |
Vacuum rate | Under no-load condition, the time for the sterilization cabinet to vacuum from atmospheric pressure to -60kpa should be ≤30min. |
Vacuum system | The use of oil-free vacuum pump, simple and convenient, fast vacuum rate, long service life. |
Multiple protection functions | With over - pressure and over - temperature protection function, more safe and reliable. |
Heating system | Adopt explosion-proof electric heating device, more convenient and fast |
Preconditioning and/or Conditioning of device through temperature & humidity variations.Gas Dwell Phase/Sterilizing Cycle where the device is exposed to the EO gas.Aeration of exposed device for removal of gas from the product.
How Does ETO/EO Sterilize?
Ethylene oxide (ETO/EO) is a colourless and odourless gas which is used as a disinfectant in hospitals and medical equipment industry for heat-sensitive tools and instruments. ETO Sterilizer is widely used in the healthcare industry due to its non-damaging effects on delicate instruments and medical devices that require sterilization thus also known as hospital sterilizer. It is used for instruments that cannot tolerate heat, moisture or abrasive chemicals, such as electronics, optical equipment, paper, rubber and plastics. The operation of EO sterilization involves placing the instruments and medical devices into a chamber for a set duration of temperature, pressure and time. Following the sterilization process, the instruments and devices are ready for use in various surgical procedures in the hospitals. The best part of using an EO Sterilizer is that it leaves no residue on items it contacts. Hospitals will continue using sterilizers because EO sterilizers have shown good sterility assurance and high throughput. Our system is specifically suitable to sterilize instruments used in CATH LABS, Medical Institutions, Eye Specialists, Ortho Specialists, General surgery Centers, Gynaec Care Units, Cardio Thoracic Surgeons, Interventional Cardiologists, ENT surgeon, Plastic / Cosmetic Surgeons, Neuro Surgeons, Gastroenterologist, Research Centers, Specialty Laboratories, Private Surgeons Nursing Homes.
EO sterilizes medical instruments and devices by alkylation. This gas infiltrates packaged medical devices to kill microorganisms and thus achieve sterilization. EO substitutes for hydrogen atoms on molecules needed to sustain life, and, by attaching to these molecules, EO stops these molecules’ normal life-supporting functions. Some of the key molecules that EO disrupts are proteins and DNA of the micro-organisms and proves lethal to microbial life.
ompliance with safety standards
EO guarantees the conformance of its machines with European standards and the ISO international regulation. As China experts in Sterilizer , we are also certified trainers which means our customers are always up to date on the latest safety rules and best practices.
In-depth knowledge of the process
SOLSTEO has a deep and detailed understanding of the ethylene oxide sterilization process. We provide our clients with training tools, a dedicated team and a network of experts, plus consulting services to keep them well briefed on topics essential to EO sterilization.
Superior Quality ETO Sterilizer Manufacturer
Sterility Equipment India Private Limited was founded in 2006. As a manufacturer and exporter of high-quality Ethylene Oxide Gas Sterilizer, Industrial ETO Sterilizer and Ribbon Blender, Sterility Equipment China Private Limited represents latest and directive technologies and incorporate an up to date know-how of modern techniques, in a combination of the most efficient customer support and service back up. We have constructed a massive infrastructural unit for streamlining our work. Our raw materials are procured from a reliable vendor base for maintaining the industry set standards.
Comments: